Onicon F-5100 Insertion Manual do Utilizador Página 20

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 33
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 19
11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax (727) 442-5699 • [email protected]
F-5100 Insertion Thermal Mass Flow Meter Manual 05/15 - 0663-20 / 18333 Page 20
INSIDE BODY VIEW
1 RED AC1 COM 1
2
S1 AC2 B + 2
3
RED NU A 3
4
WHITE NU 4
5
S2 4-20mA 5
6
WHITE SIG GND 6
A B C
24 VDC
PULSE
VDC IN
+
VDC GND
INTEGRAL
1
2
3
4
5
6
1
2
3
4
5
6
A B C
NO CUSTOMER ACCESS
DO NOT OPEN THIS SIDE
INSIDE REAR
COVER LABEL
A1 - RED - VELOCITY SENSOR WIRE (HEATED ELEMENT)
A2 - NO WIRE
A3 - RED - VELOCITY SENSOR WIRE (HEATED ELEMENT)
A4 - WHITE - TEMPERATURE SENSOR WIRE
A5 - NO WIRE
A6 - WHITE - TEMPERATURE SENSOR WIRE
B1 - AC1 - AC VOLTAGE
B2 - AC2 - AC VOLTAGE
B3 - NU - NOT USED
B4 - 4 - 20 mA POWER
B5 - VDC IN - VOLTAGE DC - POSITIVE (+)
B6 - VDC GND - VOLTAGE DC - GROUND (-)
C1 - NOT USED
C2 - RS 485 (+)
C3 - RS 485 (-)
C4 - 24 VDC PULSE - 0 TO 24 DC PULSE OUTPUT
C5 - 4 - 20 mA - 4 TO 20 mA ANALOG OUTPUT
C6 - VDC GND - VOLTAGE DC - GROUND (-)
TERMINAL BLOCK SIDE
½” NPT
USER ENTRY
FOR WIRING
(ONE ON EACH SIDE)
DISPLAY SIDE
FLOW
½” NPT
USER ENTRY
FOR WIRING
*
*
Factory installed jumper and resistor.
Do not remove.
*
3.5.3 Integral Mount Wiring
A B
C
A B
C
1
2
3
4
5
6
1
2
3
4
5
6
Control System
DC Pulse +
4-20mA +
+24 VDC
VDC Common
F-5100 Terminals
IMPORTANT NOTICE
The nominal pulse output voltage from this sensor is 24 VDC. A pulse isolator module has been
provided with this sensor that will convert this active pulse into an isolated dry contact output.
Refer to section 3.5.4 Using the Pulse Isolator Module for details.